Electrical & Photonics Engineering Lab Assistant — Quantum Opus
Plymouth, MI · May 2026 – Aug. 2026
Quantum Opus builds superconducting nanowire single-photon detectors, and the cryostats that hold them cold.
- I set up the fiber path and the attenuation, then looked at the electrical pulse.
- The detector was saturating because one laser pulse carried too many photons. Attenuating the light is how I separated that from a bad detector.
- The photo of that path is three OA 5002 attenuators. One reads 30.91 dB at 1600 nm. Two display FAIL.

Overview
I was looking at the pulse from a superconducting nanowire detector. The timing bench was a 1 GHz fiber laser, a TimeTagger, and a 500 MHz scope. The detector was saturating because one laser pulse carried too many photons. Attenuation in the fiber path, and the waveform on that scope, is what separated a light-level problem from a bad detector.
My contribution
I set up the fiber path and the attenuation, then looked at the electrical pulse. The detector was saturating because one laser pulse carried too many photons. Attenuating the light is how I separated that from a bad detector. The photo of that path is three OA 5002 attenuators. One reads 30.91 dB at 1600 nm. Two display FAIL.

Full engineering log
My work
The bench
I worked on the setups used to characterize these detectors. I prepared the fiber connections and the optical path, set the attenuation, and captured the electrical pulse. A reading only hangs together if you keep three things in view at once: the light going in, the detector and its readout, and the scope that drew the waveform.
The timing setup
That setup used a 1 GHz fiber laser, a TimeTagger for the photon arrivals, and a 500 MHz oscilloscope for the electrical waveform. 1 GHz is how often the laser pulses. 500 MHz is that scope's bandwidth. The WaveMaster in the opening photo is marked 6 GHz, so it is a different instrument.
Too many photons in one pulse
The detector was saturating because too many photons were arriving in one pulse. I found that by attenuating the light and looking at the waveform. I don't have a retest after a change. Extra photons in the pulse change the waveform you are trying to read, so a light-level problem can look like a detector problem until you check the optical path.

Tools I used
- TimeTagger
- 1 GHz fiber laser
- 500 MHz oscilloscope
- Optical attenuation
- Waveform analysis